Medfusion 3010a syringe pump
Details
The Medfusion 3010a Syringe Pump may be used in MRI rooms with static magnetic fields up to 150 Gauss. It features user-friendly, single-handed loading and a stackable horizontal design. FEATURES:•16 fusion modes •Graphic pressure trend display •Begins infusing in just 3 keystrokes •Easy to see screen •MRI Compatible •User friendly •Stackable horizontal design SPECIFICATIONS:•Dimensions: Size - 10.5" x 5.1" x 5.5"; Weight - 3.8 lbs •System Accuracy: ±2%, excluding syringe variations •Flow Rate: .03 to 1130 ml/hr, dependent upon syringe size •Delivery Modes: ml/hr; Volume/Time; Mass; Body Weight; Custom Dilution; Intermittent •Delivery Options: Volume Limit; Bolus Dose; Loading Dose; KVO Rate; Standby; Override Occlusion Limit •Delivery Libraries: Four libraries with 16 profiles each •Power: ACL 100-240 VAC, 50/60 Hz, 30Va External DC source (optional) •Recharge Time: Less 16 hours with the pump turned off •Battery Capacity: Rechargeable. 10 hours at 5.0 ml/hr with 60 ml syringe (from fully charged battery)